24/7 emergencies 814 777 6921812 230 6079WhatsApp 811 786 0768
Electrónica MonterreyElectrónica MonterreyLaboratorios Creativeit Talk to an engineer
Menu

Why Is There Ripple in a Return Loss Measurement?

Ripple appears when two or more discontinuities reflect part of the signal and those reflections add or subtract depending on frequency. A calibration with residual directivity also produces it.

The mechanism: interference between reflections

When a signal meets a discontinuity, part of it returns. If there is a second discontinuity farther along, another part returns with additional delay. At the measurement port, the two add in phase at certain frequencies and in opposition at others. The result is a periodic oscillation in the return loss trace, analogous to the interference of two waves.

This means that ripple is not noise: it has structure, and that structure contains information about the geometry of the setup.

What the ripple period tells you

The frequency spacing between two consecutive maxima is approximately the propagation velocity in the line divided by twice the distance between the two discontinuities:

frequency spacing ≈ v / (2 L)

where L is the length between the two reflecting points and v is the phase velocity in that medium. Slow ripple, with few cycles in the band, indicates discontinuities close to each other; fast ripple indicates that they are separated by a long section, such as a cable. With this relationship you can estimate where each reflection is.

Common causes

  • A poorly tightened, dirty or damaged connector and a poorly matched device at the ends of a cable.
  • Adapters between connector types, each with its own reflection.
  • A calibration with a poor load or poor standards, which leaves residual directivity; the ripple then accompanies what is being measured even if the device is fine.
  • Cables that moved after calibration.

How to interpret the result and reduce it

  1. Estimate the distance with the relationship above and look for an element in the setup that matches it.
  2. Switch to the time domain, which many network analyzers offer through a transform, to see each reflection in its position and, if appropriate, apply a time gate that isolates the device’s reflection.
  3. Eliminate unnecessary adapters and check torque and cleanliness.
  4. Add an attenuator between the device and the reflecting element to reduce the ripple, although dynamic range is lost.
  5. Recalibrate with suitable standards and at the correct plane.

If the ripple persists with a good-quality load in place of the device, the origin is in the setup or the calibration and not in the device.

We do real tests, on video! Watch them on our YouTube channel →

Do you have a problem nobody has been able to measure?

Talk directly to the engineer in charge, with no middlemen.

Talk to an engineer