Why do flexible test cables degrade VNA calibration?

Flexible test cables degrade the calibration of a vector network analyzer because, when moved, their electrical length, loss and match change, and calibration assumes those characteristics stay the same as when the standards were measured.

The assumption that breaks

Calibrating an analyzer consists of measuring known standards through the cables and calculating a correction that removes the effect of everything between the instrument and the calibration plane, cables included. That correction is valid only if the system does not change afterward. If the cables move, they change, and the calculated error no longer corresponds to the real error.

What changes when a cable is flexed

  • Electrical length and phase: bending the cable changes the geometry of the dielectric and the conductor and modifies the delay. At high frequencies, a small mechanical change is a significant phase change.
  • Insertion loss: varies with the deformation of the dielectric and the shield.
  • Match: local deformations create small reflections that add to those of the connectors and produce ripple in the response.
  • Shielding: with use, the braid wears and both leakage and sensitivity to interference increase.
  • Connectors: movement pulls on them, loosens the coupling and wears them.

What an engineer sees in practice

After calibration, the measurement of a known load shows a residual reflection that changes when the cable is moved. The trace of a device shows ripple that was not there before, or a phase shift that changes with the position of the cables. If the measurement is repeated and the results depend on how the cable is bent, the cable is the cause.

How to reduce the effect

  1. Use phase- and amplitude-stable cables, designed to withstand flexing without much change.
  2. Calibrate with the cables in the position in which they will be used and do not move them between calibration and measurement.
  3. Reduce the length and avoid tight bends; respect the manufacturer’s minimum bend radius.
  4. Secure the cables so they are not dragged by the movement of the device or the operator.
  5. Calibrate as close as possible to the device under test, or with minimal adapters.
  6. Periodically verify the calibration with a check standard and recalibrate if it deviates.

When to replace the cable

If repeatability is poor even with good practices, or there are visible crush marks, or the shield is damaged, the cable must be replaced. A degraded cable affects all the measurements that depend on it.