The cause: the meter measures total resistance
An RTD conveys temperature in the form of resistance. The meter applies a current and measures the resulting voltage, but it does not know which part of that resistance belongs to the sensor and which to the conductors. A long or thin conductor adds resistance that is interpreted as additional temperature.
Why it matters more in low-resistance sensors
The change in platinum’s resistance with temperature is small in absolute terms. With a Pt100, for example, a change of one degree equals a fraction of an ohm. As a result, a few tenths of an ohm in the wires can amount to a visibly wrong reading. In sensors with a higher nominal resistance, the same wire error carries less weight.
How each configuration deals with it
- Two-wire: the full conductor resistance goes into the result. It is acceptable only with very short cables or with a known correction.
- Three-wire: the wire resistance is measured with an additional conductor and subtracted, on the assumption that the wires have the same resistance.
- Four-wire: one pair supplies the current and the other measures the voltage, so the drop in the wires does not enter the measurement. This is the most robust option.
What the engineer sees in the field
A two-wire RTD with a long cable reads a temperature higher than the true value, and the error changes with ambient temperature because the conductors also have a thermal coefficient. With three wires, the error persists if the wires differ from one another or if oxidized connections add unequal resistance. A reading that fluctuates when the cable is moved points to a loose terminal.
How to check
Compare against a standard in a bath or block, or replace the sensor with a resistor of known value at the end of the cable. If the reading differs, the source is the wiring and not the sensor. The configuration must match that of the instrument, since setting the meter to the wrong number of wires introduces a systematic error.