Why Does a VSWR Measurement Change When the Cable Moves?

Because the cable is part of the measurement chain: flexing it changes its phase, its loss and the quality of its connectors, and a calibration made with the cable in another position no longer corrects those changes.

What really changes when the cable moves

A network analyzer calibrated at the end of a cable assumes that, after calibration, that cable does not change. Moving it alters three things. First, the electrical length and phase, because bending the cable deforms the dielectric and slightly changes the geometry. Second, the loss, which affects the signal traveling out and back when reflection is measured. Third, the contact at the connectors, if the movement pulls on them or twists them.

VSWR is calculated from the reflection coefficient measured at the calibration plane. If the cable has changed since calibration, the error terms that were corrected are no longer the real ones, and the calculated value shifts. The higher the frequency or the longer the cable, the greater the variation.

What to check first

  1. Repeatability with a load. Connect a good load to the end of the cable and move the cable. If the reflection changes, the problem is in the cable, its connectors or the calibration, not in the device under test.
  2. Transmission with a thru connection. A transmission trace that oscillates when the cable moves indicates unstable phase or loss.
  3. Connectors. They should be clean, tightened to the proper torque and free of play. An intermittent contact produces jumps.
  4. Shielding. A low-cost or damaged cable can leak or pick up signal, which changes the measurement with position and the proximity of objects.
  5. Calibration. It should have been done with the cable in the position in which it will be measured.

How to reduce the effect

  • Calibrate at the end of the cable that will connect to the device, not just at the instrument port.
  • Use cables designed for phase stability under flexing, which cost more but change less.
  • Secure the cable and do not move it between calibration and measurement; avoid small bend radii.
  • When the device’s reflection is very low, add an attenuator close to the device, because it reduces the effect of the cable’s reflections.

If the variation persists with a load and a cable in good condition, check the calibration and the torque, not the device.