Where it comes from
To measure its resistance, an RTD needs a current. That current dissipates power in the element through the Joule effect, and that power heats it above the temperature of the surrounding medium. The instrument then measures the temperature of the element, not that of the medium, and the difference is the self-heating error.
What determines its size
- The power dissipated: it grows with the square of the current and with the resistance of the element.
- Heat transfer to the medium: in stirred water, heat is removed well and the effect is small; in still air or in gas, heat is removed poorly and the effect grows.
- The sensor construction: size, contact with the sheath and internal clearance influence how much heat reaches the outside.
How it is recognized
A simple way is to measure with two different currents and compare. If the reading changes when the current is varied, there is self-heating. Another sign is the same sensor reading differently in water and in air at the same current. In a bath calibration and in a measurement in gas, the difference can be noticeable.
How it is reduced
- Use the lowest current that allows good resolution and low noise relative to the signal.
- Measure with pulsed current or limit the excitation time when the instrument allows it.
- Ensure good thermal contact between the element and the medium.
- Avoid measuring in still gas when accuracy is the goal, or improve the movement of the medium around the probe.
A detail in calibration
Certificates for standard RTDs usually state the measuring current used for calibration. Using the sensor at a different current changes the self-heating and, with it, the value. This is a key point when comparing certificates with field measurements.