Definition
NETD stands for noise-equivalent temperature difference. It is the temperature difference in the scene that generates a signal equal to the system noise. It is expressed in millikelvin: a lower value means greater sensitivity.
What it means in the image
Noise appears as a speckle that can mask subtle differences. With a low NETD, two areas with a small temperature difference can be told apart; with a high one, they merge into the noise. This is what decides whether the camera can show a faint gradient on a motor or a minimal difference between phases.
What it depends on
- Detector: technology and quality of the elements and their readout electronics.
- Optics: faster optics let more radiation reach the detector and improve the figure.
- Object and detector temperature: sensitivity is not constant across the whole range; it is specified under stated conditions.
- Integration time and averaging: more time or more averaged frames reduce noise, at the expense of speed.
What it is not
NETD describes sensitivity, not accuracy. A camera can detect very small differences and still have an absolute error of several degrees if the parameters are set incorrectly. Nor is it the same as spatial resolution, which depends on the number of pixels and the instantaneous field of view.
How to interpret the figure
- Compare figures only if they were stated under similar conditions.
- Consider sensitivity together with spatial resolution and optics.
- For inspections with large expected differences, sensitivity is no longer the limiting factor.