What is the international temperature scale?

The international temperature scale is a practical scale, defined by agreement, that assigns temperature values to reproducible fixed points and specifies how to interpolate between them with defined instruments, in order to approximate thermodynamic temperature.

Why a practical scale is needed

Thermodynamic temperature is defined from fundamental physical laws, but measuring it directly is difficult and laborious. Everyday work in laboratories and industry needs a scale that anyone can reproduce with consistent results. The international temperature scale fills that role: it is a convention that reproduces thermodynamic temperature with the best available approximation and makes it practical to realize. The current one is known as ITS-90.

What it consists of

  • Fixed points. Equilibrium states of pure substances, such as triple points and freezing points of metals, to which a temperature value is assigned. They are reproducible because they depend on the physics of the substance, not on an instrument.
  • Interpolation instruments. Between fixed points, temperature is determined with specific instruments, so that different laboratories arrive at the same value. Over much of the range of technical use, the standard platinum resistance thermometer is employed.
  • Reference and deviation functions. Equations that relate the instrument’s indication to temperature, fitted with the readings at the fixed points.

What it means for calibration

A reference thermometer is calibrated at fixed points, and working thermometers are then calibrated from it by comparison. In this way the measurement of an industrial sensor is linked to the international scale through a chain of calibrations, each with its uncertainty. That chain is traceability in temperature.

Points to remember

  • The scale is rigorously reproduced only in specialized laboratories; industrial practice uses its derivatives.
  • It is not the “true” temperature but an agreed approximation, with its own uncertainty that adds to the others.
  • The scale has been revised over time as knowledge improved; it is advisable to consult the current official definition.

What the engineer sees

On a thermometer calibration certificate, the reference to the scale indicates that the reported temperatures are expressed in it. If sensors calibrated under different scales are compared, small differences appear that are not sensor faults but differences of definition. In industrial work they rarely matter, but they do in high-level metrology.