A contribution that is not from the object
When the camera is aimed at a surface, part of the radiation it receives never came from the object: it is radiation from the surroundings that bounced off it. Reflected apparent temperature expresses that contribution as the temperature of a blackbody that would produce the same radiation.
When it carries weight
Reflection is the fraction that is not emitted, so it matters little on high-emissivity surfaces and a great deal on low-emissivity ones. On a nearly black surface, ignoring it barely changes the reading. On a shiny metal, it dominates.
How it is estimated
The reflector method is the most common:
- Place a diffuse reflector, for example a sheet of aluminum foil that has been crumpled and flattened out again, where the object is, with the crumpled side facing the camera.
- Set the camera emissivity to its maximum value.
- Measure the average temperature of the reflector: that value is the reflected apparent temperature.
- Enter that value in the camera and remove the reflector.
When the scene is a uniform environment, the air temperature is usually a good approximation.
Sources that distort it
- Nearby lamps, hot equipment, furnaces and panels.
- The sun, directly or via reflective surfaces.
- The person operating the camera, who is a heat source at the same height as the object.
What an engineer sees in practice
Hot spots shaped like the external source or like the operator. Changing the viewing angle and seeing the spot move, while a real fault stays put, is a simple test to tell a reflection from genuine heating.
If reflection dominates, it is better to change the measurement conditions (shield, change the angle, increase emissivity with tape) than to rely on the correction.