Central idea
Measuring means taking energy from the circuit. No instrument is ideal: a voltmeter draws a small current and an ammeter introduces a voltage drop. That effect on the circuit is the burden. When the instrument modifies what is being measured, the reading corresponds to the altered circuit and not the original one.
In a voltmeter
Its input impedance acts in parallel with the point being measured. If the resistance of the signal source is comparable to that impedance, it forms a divider that reduces the reading. With low-impedance sources the effect is negligible; with high-impedance circuits, such as sensors or networks of large resistors, it can be large. The solution is an instrument with higher input impedance or a buffer amplifier.
In an ammeter
It is inserted in series, and its internal resistance produces a voltage drop, sometimes called burden voltage. If the circuit is low-voltage and low-resistance, this drop reduces the actual current. Higher current ranges usually have a lower burden than low ones. A clamp meter avoids this effect because it is not inserted into the circuit.
In instrument transformers
In current and voltage transformers, the burden is the load connected to the secondary, expressed in apparent power or impedance. If it exceeds what the transformer allows, the error increases and the core may saturate. A current transformer secondary must never be left open with the primary energized, because of the risk of hazardous voltages.
What the engineer sees in practice
- A voltmeter reading changes when the instrument is replaced with one of different input impedance.
- The current in a circuit rises when the ammeter is removed.
How to estimate and reduce the effect
Compare the impedance of the instrument with that of the circuit; the ratio indicates the loading error. Choose instruments with a low burden relative to the circuit. Measurements in power installations must be made by qualified people in accordance with the applicable regulations.