What is hysteresis in a pressure sensor?

Hysteresis in a pressure sensor is the difference between the output obtained for the same pressure depending on whether it is reached going up or going down; it indicates that the sensor “remembers” where it came from.

What it is

If pressure is applied increasingly and then decreasingly, the sensor output for the same pressure may not match in the two directions. That difference is hysteresis. It is not random noise: it is repeatable and has a direction, because it depends on the sensor’s recent history.

Why it happens

  • Imperfect elasticity of the diaphragm or sensing element. Materials do not recover their shape instantly and completely after deforming.
  • Internal friction. In mechanical pressure gauges, the links, gears and Bourdon tube contribute friction.
  • Residual stresses and mounting effects. The structure, adhesive or weld retains stresses.
  • Transient thermal effects after a rapid pressure change.

How it is detected

During calibration, several points are traversed upward and then downward, and the output at each point is compared. The maximum difference between the two directions, expressed as a fraction of the span, is the hysteresis. For the measurement to be valid, you must wait for stabilization at each point and not overshoot the desired point; if you overshoot and come back, the result mixes in the effect.

What it is not

Characteristic What it describes
Hysteresis Difference depending on the direction of approach
Nonlinearity Deviation from the ideal straight line, in a single direction
Repeatability Dispersion when repeating the same approach
Drift Change over time

What the engineer sees

The typical symptom is a reading that does not return exactly to the previous value after a system is depressurized, or a control loop that settles at slightly different values depending on whether it came from above or below. On mechanical pressure gauges, a gentle tap on the dial usually reduces the friction effect, which reveals its origin.

How to interpret the result

Hysteresis that is small compared with the measurement tolerance is acceptable. Large hysteresis indicates overload, element fatigue or damage, and in most cases cannot be corrected with a zero or span adjustment.