What Is an Electrostatic Voltmeter Used For?

An electrostatic voltmeter measures voltage from the attractive force between charged electrodes, and is used to measure high DC and AC voltages while drawing practically no current.

The physical principle

An attractive force appears between two electrodes with a potential difference between them. In an electrostatic voltmeter, one of the electrodes is movable or attached to a moving element, and its displacement is translated into an indication. The force is proportional to the square of the voltage, so the indication does not depend on the polarity of the voltage.

What follows from this

  • It works on DC and on AC. On AC, because the force depends on the instantaneous square, the instrument indicates the RMS value.
  • Extremely high input impedance. It draws no current on DC and its load is only a small capacitance. It therefore does not disturb the source being measured.
  • Response over a wide frequency range, since its indication does not depend on the waveform.

What it is used for

  1. Measuring high voltage directly without a divider, when the model is designed for that range, or with the help of one.
  2. Checking high-voltage DC sources, such as those for insulation testing, where an instrument with a load current would alter the measurement.
  3. Serving as a reference for calibrating other meters, because of its accuracy and its direct relationship to a physical quantity.

When it is not the best option

Its scale is nonlinear, with more space at the upper end, and its sensitivity at low voltages is limited. It is of no use for fast impulses, since the mechanical part cannot follow a microsecond event; in those cases a divider with a recorder is used. It is also sensitive to vibration, humidity and nearby fields, so it must be placed away from objects at a different potential.

How to interpret the result

When it is used as a reference, the value read is compared with that of another measuring system; a systematic difference indicates a calibration problem in one of the two.

These activities involve hazardous voltages: they must be carried out by qualified, authorized personnel, following the applicable regulations and the facility’s lockout and grounding procedures.