How it generates force
An electrodynamic shaker works the same way as the loudspeaker in a sound system, but with more power. A coil attached to an armature sits inside a strong magnetic field. When a current flows through it, a force proportional to that current appears. The armature transmits the motion to a table on which the product or the fixture is mounted.
Control chain
A power amplifier drives the coil. A control accelerometer, placed on the table or the fixture, measures the response, and a vibration controller compares that measurement with the desired profile and corrects the drive signal. It is a closed loop: what the product receives is what the control sensor measures, not what the program requests.
What limits it
- Maximum force, which by Newton’s second law relates total mass and achievable acceleration.
- Maximum displacement, which limits the low frequencies.
- Maximum velocity, another constraint in the intermediate region.
- Resonance of the armature itself, which sets the useful upper frequency limit.
The moving mass includes the armature, fixture, product and accessories, not just the sample.
The fixture matters
The fixture must transmit the vibration without significantly amplifying or attenuating it within the test range. A fixture with its own resonance inside the band distorts the test and can overstress the sample. It is therefore designed to be rigid and light, and is verified with a response test.
Directions and axes
Many machines vibrate in a single direction. To test other axes, the sample is rotated or a horizontal slip table coupled to the shaker is used. Multi-axis systems exist, but they are a different category.
What the engineer sees
If the controller cannot achieve the profile, the cause is usually the force limit, a fixture resonance or poor accelerometer mounting. Sensor mounting, cable restraint and the tightening of the mounting bolts are part of the test. The intense magnetic fields of the armature make it necessary to keep medical implants and magnetic media away.