What Is a Vector Network Analyzer?

It is an instrument that generates a signal, applies it to a device and uses coherent receivers to measure the magnitude and phase of the reflected and transmitted waves, yielding the device's S-parameters as a function of frequency.

How it is built

A vector network analyzer (VNA) has a signal source that sweeps frequency, a set of directional couplers or bridges that separate the incident wave from the reflected wave at each port, and receivers that measure the amplitude and phase of those waves. The receivers use a common local oscillator and references that are coherent with the source. That coherence is what makes phase measurement possible.

What it measures

It measures how much of the signal applied to each port of the device is reflected and how much is transmitted: S-parameters with magnitude and phase. From these it derives quantities such as return loss, VSWR, gain, insertion loss, group delay and impedance. With suitable processing it also provides time-domain measurements.

Calibration is part of the instrument

The systematic errors of the instrument, cables and connectors are large compared with what you want to measure. The VNA corrects them with a vector calibration that uses known standards, which is why the useful data are always those measured after calibrating at the plane where the device is connected.

How it differs from a spectrum analyzer

  • The VNA generates its own signal, stimulates the device and compares the response with the stimulus. It measures the behavior of the device.
  • The spectrum analyzer only receives a signal and displays its frequency content. It measures signals.
  • The VNA measures phase and can correct errors vectorially. A spectrum analyzer with a tracking generator makes simpler scalar measurements.

Which devices are measured

Filters, amplifiers, attenuators, cables, connectors, antennas, couplers and circuits in general. Depending on the model, some analyzers also allow measurements with variable power, on mixers, or with multiple ports.

What you see in practice

Reflection and transmission curves versus frequency, with markers and limits; the Smith chart for impedance; and data exported as S-parameter files for simulation.