The physical idea
An ideal blackbody absorbs all the radiation it receives and emits the maximum radiation possible for its temperature. No real material achieves this, but it can be approximated. A reference blackbody is that approximation in instrument form: a surface or cavity whose temperature is controlled and measured with a reference contact sensor, and whose emissivity is high and known.
Cavity and plate
- Cavity type: an isothermal cavity with a small aperture. Radiation entering it is reflected many times inside and hardly escapes, so the aperture behaves like a source very close to a blackbody.
- Flat-surface (plate) type: a plate with a high-emissivity coating. It is more practical for cameras with a wide field of view, although its emissivity is somewhat lower and must be known.
What it is used for in practice
The instrument is aimed at the source, the reading is compared with the reference temperature, and the deviation is determined. Using several temperatures gives an idea of the behavior across the range of use. It is the way to find out whether a camera is still measuring correctly after a drop, a lens change, or simply the passage of time.
Conditions that affect the result
- The source must be thermally stabilized before comparing.
- The target must completely fill the instrument’s measurement area; a very small aperture in front of a low-resolution camera produces low readings due to the size-of-source effect.
- Avoid reflections from nearby hot or cold objects and air currents in front of the aperture.
- The emissivity value set on the instrument must match that of the source.
- The uncertainty of the reference (sensor, uniformity, emissivity) limits the uncertainty that can be stated for the verified instrument.
What it is not
An ice bath or a container of water does not replace a blackbody: its surface has an emissivity below one and reflects the surroundings, so the infrared reading may not match the true temperature even when that temperature is known.