The limitation of sweeping
A conventional swept analyzer observes one frequency at a time and moves across the band over time. While it looks at one region, it does not see the others. A brief signal that appears and disappears outside the observation window goes undetected. For stable signals this does not matter, but for bursts, frequency hopping or sporadic interference it is a serious problem.
How a real-time analyzer works
It digitizes a segment of the spectrum and calculates the Fourier transform of consecutive overlapping blocks, with no pauses between them. It processes at the same speed it acquires, so there is no blind time. The result is a continuous sequence of spectra across the capture bandwidth.
Key concepts
- Capture bandwidth: how much spectrum is seen at once, limited by the digitizer.
- Probability of intercept: the probability of detecting a signal of a given duration with level accuracy. It depends on the duration of the analysis block and on the overlap; a signal shorter than the minimum time is seen, but with less accurate level.
- Persistence display: it overlays many spectra with a color map according to how often each level appears. A weak sporadic signal can be seen alongside a strong continuous one, something impossible on a normal trace.
- Spectrogram: it adds time as an axis, so the history of band occupancy can be seen.
- Frequency mask trigger: it starts a capture when the signal violates a defined shape, so the event can be saved and analyzed.
What it is used for
Detecting intermittent interference, characterizing frequency hopping signals, analyzing transmitter turn-on transients and studying coexistence between systems in the same band.
When it does not apply
If only a stable carrier is being measured or a very wide frequency range is needed, a swept analyzer may be sufficient and simpler. If the goal is to demodulate and measure modulation quality, vector analysis is needed as well.