What it is for
A vector network analyzer does not measure perfectly: its directivity, the coupling between channels, the source and load mismatch and the frequency response of the signal path introduce repeatable systematic errors. Calibration determines them by measuring standards whose values are known, and then corrects subsequent measurements.
Standards it contains
- Open circuit (open): a terminal with no connection to the outer conductor. A real open has a parasitic fringing capacitance, so the kit includes it in its model.
- Short circuit (short): connects the center conductor to the outer conductor. It has a small residual inductance and a delay.
- Matched load (load): a resistance equal to the characteristic impedance, to give almost zero reflection. Loads may be fixed or sliding.
- Direct connection (thru): joins the two ports. It can be a lossless adapter or a short line of known length.
These standards are used in the most common calibration procedures, such as open, short, load and thru for two ports. Other methods, such as line-based ones, use other standards and other conditions.
Kit models
The manufacturer provides a mathematical definition of each standard, with its delay, loss and capacitance or inductance as a function of frequency. That definition must be loaded into the instrument for calibration. Using the wrong model, or the model of another kit, produces large errors that are not visible at a glance.
What to consider when choosing one
- That the connector type and frequency range match the device and the instrument.
- That the gender of the standards is right for each port: a male port is calibrated with female standards and vice versa, or else with adapters of known quality.
- That the kit model is compatible with the analyzer used.
- That it has been verified or calibrated recently, since standards wear out.
Care and checking
Standards are precision parts: they are stored clean and protected, connected with a torque wrench and careful alignment, and their contacts should not be touched. After calibrating, it is good practice to measure an independent verification standard, such as an air line or a known attenuator, to confirm that the calibration is correct.
When it does not apply
A mechanical kit calibrates up to the plane of the connectors. If there are additional fixtures or cables in the measurement, a method that includes them must be used, or they must be de-embedded afterward.