What Is a Hall Effect Current Transducer?

A Hall effect current transducer measures current from the magnetic field it produces, using a semiconductor sensor that generates a voltage proportional to that field, and it works with both alternating and direct current.

The Hall effect

When a current flows through a semiconductor material and a perpendicular magnetic field is applied to it, a transverse voltage appears between the edges of the material, proportional to the field and to the current flowing. A Hall sensor exploits this effect: with a known supply current, its output voltage is proportional to the magnetic field.

How it measures current

The current to be measured passes through a conductor that goes through a magnetic core with a small air gap. The field in the core is proportional to that current, and the sensor, placed in the gap, delivers a voltage proportional to the field. There is no electrical connection between the power circuit and the measuring circuit, so there is galvanic isolation.

Two configurations

  • Open loop: the sensor signal is amplified and delivered directly. It is simple and economical, but sensitive to temperature variations, core saturation and drift.
  • Closed loop or compensated: the output drives a secondary coil that produces an opposing field and cancels the flux in the core. The compensation current is proportional to the measured current. The sensor works close to zero flux, which gives greater accuracy, better linearity and faster response.

What sets it apart from a current transformer

A conventional current transformer measures only alternating current, because it depends on the variation of flux. The Hall transducer also measures direct current and low-frequency components, so it is used in drives, converters, chargers, traction systems and DC clamps.

Limitations

  • Zero offset or drift with temperature.
  • Remanent magnetization of the core after an overcurrent, which introduces a residual error.
  • Limited bandwidth compared with other sensors.
  • Sensitivity to external magnetic fields, from the proximity of other conductors.

How to interpret it

The output is usually a proportional voltage or current, which is converted using the manufacturer’s ratio. Transducers are installed on power conductors, work that must be done by qualified personnel.