What is a dielectric test cell with guard electrodes?

A dielectric test cell with guard electrodes is a parallel-plate sample holder in which the measuring electrode is surrounded by a guard ring at the same potential, so that the measured current passes only through the volume of the sample.

Function of the guard ring

In a simple two-plate cell, the electric field bends at the edges (edge effect) and current also flows along the surface of the sample. Both phenomena contaminate the volume measurement. The guard is an annular electrode, separated by a small gap, which surrounds the measuring electrode and is held at the same potential. In this way, the field under the measuring electrode is almost uniform and the surface current is diverted through the guard to the source, without passing through the instrument.

Components

  • Measuring electrode: defines the effective area, which in practice includes an adjustment for the gap.
  • Guard ring: connected to the shield or guard potential of the instrument.
  • High-voltage electrode or counter electrode: a larger plate that receives the voltage.
  • Clamping system: applies a reproducible, parallel force, sometimes with a micrometer to measure the thickness.
  • Support insulation: ceramic or polymer of very low conductivity.

Electrical connection

For very small currents, triaxial cabling is used: a center conductor for the signal, an intermediate shield as guard and an outer shield at ground. The intermediate shield reduces leakage from the cable itself and reduces the effect of the cabling capacitance.

Uses and variants

The cell is used for volume and surface resistivity, and for permittivity and losses at low frequency. Some cells are contact types and others use a small air gap to avoid applied electrodes, although they require a different treatment of the calculation. There are also temperature-controlled models.

What the engineer sees

If the guard is not properly connected, resistivity readings come out low and unstable, and they change when the cables are touched. Poor cleaning of the gap between electrodes produces the same symptom.

Care

  1. Keep the electrode surfaces and the gap clean.
  2. Verify the flatness and parallelism of the assembly.
  3. Measure with the cell empty to determine its background capacitance and leakage.