What each parameter measures
Capacitance indicates how much charge the capacitor stores per unit of voltage. ESR is the series resistance that represents the internal losses: those of the terminals, the electrodes and the electrolyte. A real capacitor is modeled as an ideal capacitance with a resistance in series, and that resistance determines how much it heats up and how much ripple it lets through.
Why ESR matters
In an electrolytic capacitor, aging dries out or degrades the electrolyte. ESR increases while the capacitance may stay close to its initial value. In a switching power supply, ripple current flows through the capacitor; with high ESR, it dissipates more power, heats up and gets worse, and the output voltage shows more ripple. That is why a component with acceptable capacitance but high ESR can be the cause of a failure.
How the instrument works
It applies a small-amplitude AC signal, at a frequency high enough for the capacitive reactance to be negligible compared with the resistance, and measures the resistive part of the impedance. Many models have a low test voltage level so they can measure in circuit, although always with reservations.
How to interpret the reading
- Discharge the capacitor and disconnect power from the equipment.
- Measure capacitance and ESR; if possible, out of circuit.
- Compare the ESR with the datasheet or with a new capacitor of the same type.
- Keep in mind that ESR changes with temperature and frequency, so comparisons must be made under similar conditions.
What to check first
An ESR clearly above what is expected for that type and size indicates a degraded capacitor, even if its capacitance is normal. It is also advisable to check neighboring capacitors from the same batch or those working under the same thermal conditions, because they age in a similar way.
Capacitors can retain a dangerous charge; the work must be carried out by qualified personnel in accordance with applicable regulations.