General scope
CISPR 16 is a series of specifications for radio disturbance and immunity measuring instruments and methods. Part 1-4 deals with ancillary equipment for radiated disturbances: among other aspects, the antennas and test sites used in radiated emissions measurements. Other parts of the series cover measuring receivers, coupling networks and test methods.
Why measurement antennas are standardized
A radiated emissions test compares the measured field with a limit. If each laboratory used an antenna with a different response, or a site with different reflections, the results would not be comparable. That is why the standard establishes what characteristics antennas must have and how the suitability of the site is verified.
What kind of requirements it addresses
- Characteristics of measurement antennas, such as their antenna factor and calibration.
- Conditions of test sites, such as open-area sites or semi-anechoic chambers, and the procedures for validating them.
- Quality criteria for the environment, so that reflections do not distort the measurement.
What the engineer sees in practice
The laboratory uses calibrated antennas with a current antenna factor and periodically validates the site. Broadband antennas, such as biconical or log-periodic antennas, are often used for their wide coverage. The antenna factor is applied to the receiver reading to obtain the field strength.
How to interpret it
- Determine which part of the series applies: receivers, antennas and sites, or methods.
- Confirm that the antenna and site meet what is required by the product standard that sets the limits.
- Consult the current edition, because the series is revised and amended regularly.
What it does not define
CISPR 16-1-4 does not set the emission limits for a product. Those limits come from other standards, which refer to CISPR 16 for the measuring equipment and method.