What are S-parameters?

They are the coefficients that describe how a device with one or more ports reflects and transmits incident waves; each parameter is the complex ratio between a wave leaving one port and a wave entering another.

The basic idea

At RF and microwave frequencies it is difficult to measure voltages and currents directly, and the open circuits and short circuits that other parameters require are impractical or make active devices oscillate. S-parameters avoid this by describing the device in terms of waves: what arrives at each port, what is reflected and what leaves through the others.

How they are defined

Each port has an incident wave (entering the device) and an outgoing wave (leaving it), normalized to a reference impedance. The S-parameter with subscripts ij is the ratio of the outgoing wave at port i to the incident wave at port j, with all other ports terminated in the reference impedance so that they reflect nothing. Each parameter is a complex number: it has magnitude and phase.

How to read them on a two-port device

  • S11. Reflection at port 1 with port 2 terminated. It indicates how well matched the input is.
  • S21. Transmission from port 1 to port 2. In an amplifier it represents gain; in a filter or cable, insertion loss.
  • S12. Reverse transmission, related to isolation.
  • S22. Reflection at the output.

They are usually expressed in decibels for magnitude and degrees for phase, or as points on a Smith chart for reflections.

What they assume and what limits them

  • The reference impedance must be the same one with which the measurement was calibrated; changing it changes the values.
  • They describe linear behavior. An amplifier driven by a strong signal stops being linear, and its small-signal S-parameters no longer describe its output.
  • A set of parameters is valid for a given set of bias, temperature and frequency conditions.

What they are used for

They make it possible to predict how a cascade of devices will behave, calculate matching, evaluate amplifier stability and compare measurements with simulations. They are normally exchanged in text files with tables by frequency. The quality of the data depends on the calibration of the analyzer that measured them.