What are interchangeable lenses in thermography?

They are optics of different focal lengths mounted on the same thermal imaging camera to change the field of view and the minimum object size that can be measured accurately.

What changes when you change the lens

A thermal imaging camera forms the image with a detector of a certain number of pixels. The lens decides how much of the scene that detector covers. A wide-angle lens takes in a lot, and each pixel “sees” a large area; a telephoto lens takes in little, and each pixel sees a small area. The quantity that describes this is the instantaneous field of view (IFOV): the angle subtended by a single pixel.

Long-wave infrared optics are not made of ordinary glass, which is opaque in that region, but of materials such as germanium, which is why they are expensive and delicate.

Why it matters for measurement

A pixel reports an object’s temperature correctly only if the object fills it completely. If the target is smaller than the area covered by a pixel, the reading mixes the object with the background and underestimates the temperature. So when inspecting a small connector from several meters away, a telephoto lens improves the measurement without moving closer to the equipment; a wide-angle lens, on the other hand, is useful inside an electrical room or in front of a large switchboard, where there is little space to step back.

What the engineer sees in practice

  • With the wide-angle lens from far away, the hot spot appears small and with a low apparent temperature.
  • With the telephoto lens, the same anomaly is resolved in more detail and the temperature rises toward its true value.
  • With the telephoto lens, focus is more critical and any vibration or hand movement is more noticeable.

What to watch for when changing lenses

  • The camera must recognize the installed lens, because the radiometric calibration includes the effect of the optics. A lens not paired with that camera body can give inaccurate readings.
  • Mount and remove the lens with the equipment switched off, and avoid touching the optical surface, which has an anti-reflective coating.
  • Keep dust off the detector during the change.
  • Record in the report which lens was used, because it changes the distance from which a measurement is valid.

A selection criterion

The lens is chosen according to the smallest target that matters and the safe working distance. When the equipment is energized, that distance is set by the applicable safety procedure and is observed by qualified personnel; the right lens makes it possible to measure well without shortening it.