Definition
A test profile is the description, as a function of time, of the conditions applied to the sample. It is like a recipe: it states which variable changes, to what value, how quickly, how long it is held and how many times it is repeated. Two tests with the same extreme values but different profiles can produce very different results.
Elements of a profile
- Levels: the extreme or plateau values of each variable.
- Ramps: the rate of change between levels, which determines the thermomechanical stress.
- Dwells: the time each level is held, which must allow the sample to stabilize.
- Cycles: the repetitions, which accumulate fatigue.
- Sample state: on, off or operating, and the points at which its operation is verified.
Why the shape matters
A fast ramp produces gradients between the interior and exterior of the sample, and stresses that a slow ramp does not generate. A short dwell does not allow the thermal mass of the product to reach equilibrium, and the test ends before the sample reaches the intended temperature. It is common to confuse the chamber temperature with the sample temperature, which is why both are measured.
What the engineer sees
The profile is represented on a graph with time on the horizontal axis and the variables on the vertical axis, and it is compared with the actual record. Differences between what was programmed and what was measured, such as overshoot or sample lag, are important data for interpreting the test.
Where the profile comes from
It can come from a test standard, from the customer’s specification or from an analysis of the use environment. Values must be taken from those sources, not assumed. A poorly defined profile produces a test that is neither demanding nor representative, even if the chamber works perfectly.