The problem it solves
A confined space is not a homogeneous mixture. Gases heavier than air, such as the vapors of many solvents, hydrogen sulfide, argon or carbon dioxide, collect at the bottom. Lighter ones, such as methane or hydrogen, concentrate at the top. If you measure only at the access opening, you get the reading for the best-ventilated spot, not for the place where the person will be.
How it is done
- Plan the points. At a minimum the top, middle and bottom zones, and along the length if the space is horizontal. In spaces with partitions, levels or corners, each one is sampled.
- Use a probe and hose to measure from outside without anyone entering. The hose is lowered slowly.
- Wait at each point. The sample needs time to reach the sensor through the hose, and the sensor needs time to respond. If the hose is lowered too quickly, the reading corresponds to an earlier point.
- Record each point with its value and location.
- Repeat after ventilating and whenever the activity changes.
Practical details
- Long hoses increase the delay and can adsorb certain gases; follow the lengths recommended by the manufacturer.
- The hose must be intact, with no kinks or loose connections, so that outside air does not get in.
- The probe must not be immersed in liquids or sludge, to avoid damaging the instrument.
- The reading order does not change: oxygen, combustibles and toxics.
When there are obstructions
Baffles, coils or piping can create pockets where gas is trapped. In these cases the probe is brought as close as possible to each zone, or the geometry is inspected beforehand. Sludge-covered bottoms deserve special attention because they can release gas when disturbed.
What the result indicates
If there are marked differences between levels, the space is stratified and ventilation is not mixing the air well; the flow must be improved or the low zone treated as the governing condition. The reading from the worst-case zone is the one that decides. This measurement and the entry must be carried out by trained people, in accordance with current regulations.
This guide is for information only. Work involving electrical hazards, heights or hazardous atmospheres must be carried out by qualified, authorized personnel in accordance with applicable regulations.