How is an impact test performed to find natural frequencies?

The structure is struck with an instrumented hammer while an accelerometer records the response; the ratio of response to force, as a function of frequency, shows peaks at the natural frequencies.

Principle

A short blow contains energy over a wide frequency range. The structure responds more strongly at its natural frequencies. If the force of the blow and the response are measured, their ratio in the frequency domain (the frequency response function) shows peaks at those frequencies, and the phase changes when passing through each one.

Equipment

  • Hammer with a force sensor in the head.
  • Accelerometer attached to the structure at the point of interest.
  • Two-channel analyzer or acquisition system with FFT calculation.
  • Interchangeable impact tip.

Procedure

  1. Stop the machine and lock it out, in order to measure the structure at rest; the test can also be done with the machine running using specific techniques, but it then mixes with the operating vibration.
  2. Attach the accelerometer at the point and in the direction to be evaluated.
  3. Choose the tip: a harder tip excites a wider frequency range; a softer one concentrates the energy at low frequencies.
  4. Strike cleanly, with a single impact per capture, in the direction of the sensor.
  5. Average several impacts to reduce noise.
  6. Check the coherence between force and response.

What to watch for

  • Avoid double hits and rebounds, which distort the force spectrum.
  • Make sure the blow does not saturate the channels and that the record covers the full decay of the response.
  • Verify that the force has enough content in the band of interest.
  • Repeat at several points, because a structure has several modes and one may not be excited from a given point.

How to interpret the result

The peaks in the spectrum are natural frequency candidates, with high coherence and a phase change at each one. They are compared with the machine’s excitation frequencies, such as the rotating speed and its harmonics, to see whether there is a risk of resonance. A wide margin between the two is desirable. If they coincide, changes in stiffness, mass or speed are evaluated.