How is water in oil measured with coulometric Karl Fischer?

In coulometric Karl Fischer, the water in the sample reacts with iodine generated electrolytically inside the cell, and the amount of water is calculated from the electric charge consumed, with no need for a prepared titrant. It is the suitable method for very low contents such as those in insulating oil.

Principle

The Karl Fischer reaction consumes iodine and water in a fixed proportion in the presence of sulfur dioxide, a base and an alcohol. In the coulometric method, iodine is produced at the anode by oxidation of iodide, using a controlled current. Faraday’s law relates the electric charge to the amount of iodine produced, and that amount to the amount of water. The charge required until a detector indicates that no water remains is measured; the charge gives the mass of water directly.

Why it is preferred for oil

Insulating oil contains water in very small amounts. The volumetric method loses accuracy at those levels, while the coulometric method is very sensitive and does not require calibrating a titrant, because it is based on a fundamental constant.

General procedure

  1. Condition the cell until a low, stable drift is reached, which represents the ingress of ambient moisture.
  2. Take a known amount of oil with a dry syringe, preferably determined by weighing the syringe before and after injection.
  3. Inject the sample through the septum, without touching the reagent with the needle, and record the charge up to the end point.
  4. Calculate the water content as mass of water divided by mass of oil, correcting for drift.

Two ways of introducing the sample

  • Direct injection: simple, but oil is poorly miscible with the reagents, so a suitable solvent is sometimes added or a special reagent is used to improve water extraction.
  • Evaporation oven: the sample is heated and a dry gas carries the water vapor to the cell. This keeps the oil from fouling the cell.

Sources of error

  • Ingress of ambient moisture when handling the sample or the syringe.
  • A non-homogeneous sample, or free water settled at the bottom of the container.
  • High drift that is not corrected.
  • Depleted or contaminated reagent, or a saturated diaphragm.
  • Substances that react with iodine and produce false readings; this is uncommon in mineral oils, but can occur with additives or contaminants.

What an engineer sees in practice

Results that fluctuate between injections of the same sample are usually due to handling or heterogeneity, not to a problem with the instrument. To interpret the value, record the sampling temperature and compare it with the solubility of the oil, as is done with relative saturation.