How is vacuum pressure measured?

Vacuum pressure is measured with the sensor suited to the pressure range: diaphragm gauges in low and medium vacuum, thermal gauges in an intermediate range and ionization gauges in high vacuum; many of them depend on the type of gas.

No single gauge does it all

Vacuum pressure spans many orders of magnitude. A single physical principle cannot cover all of it, which is why different gauges are chosen according to the range.

Main types

Type Principle Characteristic
Mechanical and capacitance diaphragm Deflection of a diaphragm by the pressure difference Measure true pressure, regardless of the gas; useful in low and medium vacuum
Thermal (Pirani, thermocouple) Heat loss from a filament, which depends on the thermal conductivity of the gas Reading differs with the gas; require a correction factor
Ionization (hot or cold cathode) The gas is ionized and the current is measured For high vacuum; sensitive to the gas and to contamination

Gas dependence

Thermal and ionization gauges do not measure pressure directly, but rather a property of the gas related to it. An instrument calibrated for air or nitrogen gives a different reading if the gas is something else. Diaphragm gauges, by contrast, respond to force and do not depend on composition.

Installing the port

  • Place the sensor near the point of interest, with a short duct of good diameter, because at low pressure conductance limits equilibrium.
  • Avoid ports near the pump or near vapor sources that disturb the reading.
  • Pay attention to cleanliness: oil or particle contamination alters ionization and thermal gauges.
  • Allow time for the walls to outgas.

What the engineer sees

Different readings between two gauges on the same system are usually due to gas dependence, different port locations or contamination. A reading that falls slowly may reflect outgassing and not a bad gauge.

How to interpret the result

You must always know the gauge type, the gas and the range over which the gauge is reliable. A value at the end of a sensor’s range is less reliable than one in its central zone.