What relationship is measured
Permeability relates the magnetic induction to the field that produces it. In linear materials and at low amplitude it is defined as a relative permeability, a dimensionless number. At high frequencies it becomes complex: one part describes the storage of magnetic energy and the other the losses.
The toroid method at low frequency
A ring of the material is made, wound with one or several turns, and the inductance is measured with an impedance meter. Because the flux is confined to the ring, there is almost no leakage and the permeability is calculated from the inductance, the number of turns and the dimensions. A second winding makes it possible to measure the induced voltage and obtain the magnetization curve.
Radio frequency and microwaves
- Toroid fixtures: a sample holder that fits the ring around a conductor and connects to an impedance analyzer, useful up to high radio frequencies.
- Coaxial line or waveguide: the sample fills a section and scattering parameters are measured; permittivity and permeability are extracted from them at the same time.
- Resonant methods: at one frequency, with good sensitivity.
What to bear in mind
- Permeability depends on field amplitude, frequency and temperature, so every result must state those conditions.
- Materials saturate: above a certain induction, the incremental permeability falls.
- Magnetic history matters; the sample is usually demagnetized before measurement.
- With a poorly controlled air gap in the setup, the apparent permeability drops.
What the engineer sees
In many ferrite materials, permeability stays almost constant up to a certain frequency and then falls, while magnetic losses increase. That behavior defines the use of the material: filters, transformer cores or absorbers.
How to interpret the result
Permeability measured at small signal does not predict behavior at high current. Power applications require curves at the real operating amplitude.