How is temperature measured with a thermocouple?

Temperature is measured with a thermocouple by connecting it, with the correct cable type and polarity, to an instrument that compensates the reference junction and converts voltage to temperature; quality depends above all on the connections and the installation.

The measurement chain

Measuring with a thermocouple involves several elements: the sensor, the cable that connects it to the instrument, and the instrument with its reference junction compensation and voltage-to-temperature conversion. An error in any link is reflected in the reading.

Steps for measuring well

  1. Choose the type. Each thermocouple type has its own range, behavior and resistance to different atmospheres. The instrument must be configured for the same type.
  2. Choose the construction. Exposed, insulated or grounded junction; with or without a sheath. An exposed junction responds quickly but becomes contaminated; an insulated one avoids ground loops; a grounded one responds faster than an insulated one but can introduce noise.
  3. Use cable and connectors of the corresponding type. Extension and compensating cables are specific to each type. Copper cable is a poor substitute on any section and creates a new thermocouple.
  4. Respect the polarity. Reversing it gives incorrect readings that sometimes look reasonable.
  5. Compensate the reference junction. The instrument normally does this with an internal sensor, so its terminals must be at a uniform temperature.
  6. Install with sufficient immersion. The sensor must have thermal contact with what it measures and enough immersed length that heat conducted along the sheath does not distort the reading.
  7. Protect against noise. The signal is on the order of millivolts; twisted, shielded cable, kept away from power cables, prevents interference.

Phenomena that affect accuracy

  • Lack of wire homogeneity in zones with a temperature gradient, which generates stray voltages.
  • Aging and contamination of the wire at high temperature, which change its response over time.
  • Insulation leakage at very high temperature.
  • Heat conduction along the sheath when immersion is short.

Common errors

  • Configuring the instrument for one type and connecting another.
  • Splicing with ordinary cable.
  • Placing the reference junction next to a heat source.

What to check first

If a reading looks wrong, check the configured type, polarity, cable and terminal temperature before suspecting the sensor. A reading of ambient temperature when the sensor is at a hot spot usually indicates an open cable or a disconnected sensor.