The measurement chain
Measuring with a thermocouple involves several elements: the sensor, the cable that connects it to the instrument, and the instrument with its reference junction compensation and voltage-to-temperature conversion. An error in any link is reflected in the reading.
Steps for measuring well
- Choose the type. Each thermocouple type has its own range, behavior and resistance to different atmospheres. The instrument must be configured for the same type.
- Choose the construction. Exposed, insulated or grounded junction; with or without a sheath. An exposed junction responds quickly but becomes contaminated; an insulated one avoids ground loops; a grounded one responds faster than an insulated one but can introduce noise.
- Use cable and connectors of the corresponding type. Extension and compensating cables are specific to each type. Copper cable is a poor substitute on any section and creates a new thermocouple.
- Respect the polarity. Reversing it gives incorrect readings that sometimes look reasonable.
- Compensate the reference junction. The instrument normally does this with an internal sensor, so its terminals must be at a uniform temperature.
- Install with sufficient immersion. The sensor must have thermal contact with what it measures and enough immersed length that heat conducted along the sheath does not distort the reading.
- Protect against noise. The signal is on the order of millivolts; twisted, shielded cable, kept away from power cables, prevents interference.
Phenomena that affect accuracy
- Lack of wire homogeneity in zones with a temperature gradient, which generates stray voltages.
- Aging and contamination of the wire at high temperature, which change its response over time.
- Insulation leakage at very high temperature.
- Heat conduction along the sheath when immersion is short.
Common errors
- Configuring the instrument for one type and connecting another.
- Splicing with ordinary cable.
- Placing the reference junction next to a heat source.
What to check first
If a reading looks wrong, check the configured type, polarity, cable and terminal temperature before suspecting the sensor. A reading of ambient temperature when the sensor is at a hot spot usually indicates an open cable or a disconnected sensor.