How is leakage current measured in portable equipment?

Leakage current in portable equipment is measured with the equipment connected and running under controlled conditions, using an instrument that records the current flowing to ground or to accessible parts, to confirm that it poses no risk to the user.

What is being measured

All electrical equipment has a small current that diverts to ground or to its accessible parts through the capacitance of filters and insulation and through the finite conductivity of materials. If that current grows, the insulation is deteriorating or there is a fault. In portable equipment, which is held in the hand, it matters especially because the user can become part of the path.

Types of reading

  • Protective conductor current: the current flowing through the equipment’s grounding conductor.
  • Touch current or accessible-part current: the current that would flow through a person touching the housing, measured with a network that simulates the impedance of the body.
  • Differential leakage current: the vector difference between the outgoing and the return current.

How it is done

  1. Inspect the equipment, the cord and the plug before the test.
  2. Check the continuity of the protective conductor, if the equipment has one.
  3. Connect the equipment to an electrical safety tester, or use a leakage clamp around the line and neutral conductors together (for differential leakage) or around the protective conductor (for its current).
  4. Operate the equipment under normal conditions and, depending on the method, with normal and reversed polarity and with the switch in each position.
  5. Record the highest reading and compare it with the acceptance criterion in the product standard or the company procedure.

Factors that alter the result

  • Moisture and dirt on the insulation surface.
  • Interference filters, which contribute normal, legitimate leakage.
  • Waveform: equipment with power electronics produces current with high-frequency content that the instrument must measure correctly.
  • Temperature of the equipment in operation.

How to interpret the result

A reading within the criterion indicates acceptable insulation. An upward trend between successive tests is a sign of deterioration even if the limit has not yet been exceeded. A high result means removing the equipment from service until it has been reviewed.

This test involves operating energized equipment and must be carried out by qualified personnel following the applicable regulations. Consult the current criterion in the product standard.