What is really being measured
A combustible gas detector does not count molecules of a particular gas: it estimates how close the mixture is to being able to ignite. The scale is the percentage of the lower explosive limit of the gas used to calibrate the sensor. If the actual gas is different, the reading is interpreted with a correction factor supplied by the manufacturer.
Sensor technologies
- Catalytic: the gas oxidizes on a hot element; the temperature change is converted into a signal. It requires oxygen.
- Infrared: it measures the absorption of infrared light by hydrocarbons. It works even without oxygen.
- Thermal conductivity: it measures changes in the gas’s ability to conduct heat; it is used for higher concentrations and for certain gases.
Procedure
- Functional test of the instrument with reference gas.
- Measure oxygen first, because some sensors depend on it.
- Sample from outside with a probe, waiting for the reading to stabilize.
- Sample at several levels: vapors from liquid fuels are usually heavier than air and collect low down; methane and hydrogen are lighter and rise.
- Keep measuring continuously for as long as work is under way.
Factors that distort the reading
- Catalytic sensor poisoning by silicon compounds, certain halogenated compounds or sulfur, which reduce its sensitivity with no obvious warning. That is why verification with gas is essential.
- Gas different from the calibration gas: the reading may end up below or above the true value.
- Oxygen-free atmosphere or excess fuel: the catalytic sensor may give low or confusing readings.
- Dusts: gas detectors do not detect combustible dusts.
How to interpret the result
A zero reading does not prove the absence of danger if the sensor is not suitable for that gas. Any sustained reading above zero calls for finding the source and ventilating. Alarms are set well away from the ignition limit. On a rapid rise, get out and investigate from outside. Measurement and entry decisions are the responsibility of trained personnel, in accordance with applicable regulations.
This guide is for information only. Work involving electrical hazards, heights or hazardous atmospheres must be carried out by qualified, authorized personnel in accordance with applicable regulations.