How is carbon monoxide measured in a workspace?

Carbon monoxide is measured with a specific electrochemical sensor, in the breathing zone and near combustion sources, and is interpreted by concentration and exposure time, because it is colorless, odorless and binds to hemoglobin, preventing oxygen transport.

Why an instrument is needed

Carbon monoxide (CO) cannot be seen or smelled. It is produced by incomplete combustion: gasoline or diesel engines, heaters, furnaces, welding, fires. When inhaled, it binds to hemoglobin with much greater affinity than oxygen and reduces the blood’s capacity to transport oxygen. The initial symptoms, such as headache, dizziness or nausea, are nonspecific and are mistaken for tiredness.

How it is measured

Most portable detectors use an electrochemical sensor: CO oxidizes at an electrode and produces a current proportional to its concentration. These sensors are selective, relatively fast to respond and suitable for personal use. An alternative is field analyzers using another technology, for spot measurements.

Where and how to measure

  • In the breathing zone of the exposed person, which is what determines the dose.
  • Near the sources to locate the origin: exhausts, burners, poorly ventilated areas.
  • During the activity, not just at the start, because the concentration varies with the operation of the equipment.
  • In spaces with engines, even if they are near the entrance, since exhaust gases can get in.

Because the density of CO is very close to that of air, it tends to mix through the space rather than stratify markedly, but hot air from combustion can rise and collect at ceilings.

Sensor limitations

  • Cross-sensitivity: some sensors also respond to hydrogen or other gases, which can give high readings with no CO present.
  • Limited service life and drift over time; functional tests and calibration are required.
  • Extreme humidity and temperature affect the response.

How to interpret

The effect of CO depends on the concentration and on how long it is breathed. That is why exposure criteria consider an average over the shift as well as a limit for short periods; the specific values are set by the applicable regulations, which must be consulted. If the reading rises, the source is eliminated or ventilated; if the alarm sounds, get out. Avoiding the use of combustion engines inside enclosed spaces is a more effective measure than measuring. The assessment must be done by trained personnel.

This guide is for information only. Work involving electrical hazards, heights or hazardous atmospheres must be carried out by qualified, authorized personnel in accordance with applicable regulations.