How is the input impedance of an antenna measured?

Input impedance is measured by connecting the antenna to a vector network analyzer, previously calibrated at the connector plane, and reading the complex reflection coefficient as a function of frequency, from which the impedance is calculated.

What is actually measured

A vector network analyzer does not measure impedance directly. It measures the complex reflection coefficient (magnitude and phase) at its port and, using the system reference impedance, calculates the input impedance. Because the result is complex, it has a real part (resistance) and an imaginary part (reactance). Zero reactance marks the resonance, and the real part indicates how close the antenna is to the impedance of the line.

Procedure

  1. Calibrate the analyzer at the end of the test cable, with the open, short and load standards, so that the reference plane lies at the antenna connector.
  2. Connect the antenna in its real configuration of use, with a representative support, ground plane and environment.
  3. Sweep the frequency range of interest and record the reflection coefficient.
  4. Present the result as impedance, return loss or a Smith chart, depending on the analysis required.

What the engineer sees on the screen

On the Smith chart, a well-matched antenna traces a curve that passes close to the center. A resonant antenna crosses the horizontal axis. If the curve shifts when the test cable is turned or a hand is brought close, the environment is influencing the measurement.

The sources of error that weigh most

  • Misplaced reference plane: the cable length rotates the phase and makes it look as if the reactance has changed.
  • Currents on the cable shield: with balanced antennas, the test cable becomes part of the antenna. A balun or a ferrite choke reduces the effect.
  • Nearby objects: metal tables, people and walls alter the impedance.
  • Cables flexed after calibration: these change the system response.

How to interpret the result

The input impedance of an antenna depends on frequency and on the environment, so a single value does not describe it. What is useful is the complete curve and the bandwidth over which the match is acceptable for the application. If the measured impedance differs from that of the line, a matching network can be added, but it is advisable to first confirm that the difference does not come from the measurement setup.