What needs to be known
Each time a connector is mated, the exact position of the contacts and the pressure vary slightly. That variation modifies reflection and phase, and it is therefore a component of the uncertainty of any measurement made with connectors. Repeatability quantifies how much the result changes between successive connections made under the same conditions.
General procedure
- Calibrate the vector network analyzer at the reference plane of the connectors.
- Connect the device or standard (for example, a load or a line) with a torque wrench set to the manufacturer’s value, letting the nut turn, not the connector body.
- Measure the S-parameters, disconnect, and repeat the cycle many times, without recalibrating between cycles.
- Compare the measurements and calculate, at each frequency, the deviation between them, either as a standard deviation or as a maximum difference.
- Present the spread as a function of frequency, for magnitude and for phase, or as an equivalent reflection.
What influences the result
- Torque: an inconsistent value changes the contact pressure. A correct torque wrench reduces the variation.
- Alignment and cleanliness: misaligned pins, dust or residue increase the spread.
- Wear and damage: connectors with many cycles or scratches have poorer repeatability. Pin dimensions, checked with a gauge, reveal wear.
- Connector type: precision connectors with an air dielectric and careful coupling are more repeatable than general-purpose ones.
- Cable movement: if cables move during the tests, this effect is mixed with that of the connector. It is advisable to secure or isolate them.
- Orientation when connecting: rotating the connector between cycles reveals asymmetries.
How to interpret the result
A spread that is small compared with the target uncertainty indicates that the connector does not limit the measurement. A spread that grows with frequency is expected, because mechanical differences weigh more as the wavelength gets shorter. If the spread is excessive, the condition of the connectors must be checked before blaming the instrument.
An example
When ten connections of a load are repeated and the reflection is seen to vary much more in one of the orientations, a damaged pin is suspected rather than an instrument problem.