How is the emissivity of a surface determined in the field?

Emissivity is determined in the field by comparing the camera reading with a known temperature of the same surface, or by applying a material of known emissivity and adjusting until the readings match.

Why it is done in the field

Tables give general values, but actual emissivity depends on the finish, oxidation, dirt and the condition of the paint. When the decision matters, it is determined on the surface itself.

Reference temperature method

  1. First set the reflected apparent temperature and the distance, because they affect the calculation.
  2. Measure the actual surface temperature with a reliable contact sensor, ensuring good thermal coupling and that the surface is at steady state.
  3. Aim the camera at a point close to the contact point, at the same angle that will be used later for measurement.
  4. Change the emissivity in the camera until the reading matches that of the sensor.
  5. Record that value for that surface.

Method with a material of known emissivity

A tape or matte paint with a known, high emissivity is placed on the surface. It is allowed to reach the surface temperature, then measured using the emissivity of the coating, and the temperature is recorded. The camera is then aimed at the bare surface next to the tape, and the emissivity is adjusted until the same reading is obtained. This method is useful on low-emissivity metals.

How to interpret the result

  • If the value obtained is very low, the measurement will be sensitive to reflections, and it is advisable to improve the surface rather than rely on the correction.
  • The result is valid for that material, finish, temperature and angle; it must not be extrapolated to other surfaces.
  • A difference between two points on the same component may be a variation in emissivity and not in temperature.

When it does not apply

On energized equipment, no sensor or coating is applied without the procedures and qualifications required by applicable regulations. In that case, the choice is to use conservative values and present the result as qualitative or comparative.