How is the calibration point of a sensor chosen?

The calibration point is chosen according to how the sensor will be used: the points must cover the actual working range, include the most critical process value and be enough to reveal the sensor's behavior, which depends on whether it is linear or not.

Guiding criterion

Calibration must answer this question: how well will this sensor measure where I will actually use it? The points are chosen to give that answer, not out of habit or for the convenience of the equipment.

Factors that determine the choice

Factor Effect on the choice
Actual range of use The points must cover at least the full span of what the process demands, not necessarily the sensor’s maximum range
Critical process point If you work around a control value, it is advisable to include it or concentrate points around it
Sensor linearity A nearly linear sensor needs fewer points; one with marked curvature requires more to describe its shape
Required tolerance The tighter it is, the more points are needed, and the better distributed they must be, to support the conclusion
Sensor history If it has drifted more at some part of the range, that part deserves attention
Capability of the calibration medium Baths and furnaces have temperature and stability limits; pushing them to the maximum reduces quality

Sensor types

  • Platinum resistance sensors: their behavior is smooth and well modeled; a few well-distributed points describe their curve.
  • Thermocouples: their relationship is more irregular and, with aging, the deviation can vary with temperature; more points are justified, especially at high temperatures.
  • Pressure sensors: they must include ascending and descending points to reveal hysteresis, in addition to zero and the upper end.

An example

A temperature sensor capable of a wide range is used in a process that always operates in a narrow zone. Calibrating it over its full span generates information about zones that are not used and leaves less resolution where it matters. It is better to concentrate the points in the working zone and add one or two outside it as a margin.

When it does not apply

If a standard, specification or customer requirement defines the points, those points are followed; the logic above serves to decide on additional points or when the criterion is open.

How to interpret the result

A good set of points makes it possible to conclude whether the sensor is fit for use and, if it no longer is, in which part of the range. A poorly chosen set can show a “good” instrument precisely where it is not used and hide the defect where it is.