How is the power of a microwave signal measured?

With a power meter and a sensor suited to the frequency, level and type of signal, applying the calibration factor for the frequency and taking into account the match between source and sensor.

Why a sensor is used instead of an analyzer

A spectrum analyzer is frequency-selective, but its absolute level accuracy is limited. A power meter with a sensor measures the total power reaching the sensor, with greater accuracy and traceability. If there is only one signal, it is the most accurate way to measure its absolute power.

Sensor types

  • Thermal (thermocouple, thermistor). They convert the absorbed power into heat and measure the temperature rise. They respond to true average power, regardless of the waveform. They are ideal for modulated signals and signals of complex amplitude.
  • Diode. They detect the signal with a diode. They reach lower levels and respond quickly, which suits them to pulses, but their response stops being square-law at high levels, and wideband models correct for this.

Measurement steps

  1. Choose a sensor whose frequency and power range covers the signal, without exceeding the maximum permitted level.
  2. Zero the meter with no signal present, to remove the offset.
  3. Connect using clean connectors and the correct torque.
  4. Enter the frequency so that the meter applies the sensor’s calibration factor at that point.
  5. Read the power, allowing the averaging time needed for the reading to stabilize.

Sources of error

  • Mismatch. If the source and sensor are not perfectly matched, part of the power is reflected and there is uncertainty due to interference between the reflections. A well-matched attenuator or an isolator between them reduces it.
  • Sensor at its limit. A level too close to the sensor’s floor or maximum degrades accuracy.
  • Average and peak power. With pulses, average power is lower than peak power; they are related by the duty cycle, and peak measurements require specific sensors and meters.
  • Harmonics. The sensor measures everything that reaches it, including the harmonics of the source.

How to interpret the result

The reading is total absolute power, in watts or dBm. For high powers, a calibrated attenuator is inserted and its attenuation is added to the reading.