What a counter measures
A frequency counter counts the cycles of the signal during a defined time interval, called the gate, which it generates from its own time base. The frequency is the number of cycles divided by the gate duration. The accuracy of the measurement therefore cannot exceed the accuracy of that time base.
Accuracy and resolution are not the same
- Resolution. The smallest change the counter can distinguish. It improves with a longer gate, because more cycles are counted.
- Accuracy. How close the reading is to the true frequency. It depends on the time base, its adjustment, aging and temperature.
You can have many digits of resolution on a misadjusted time base, and the figures will be misleading.
How to achieve an accurate measurement
- Use an instrument with a time base of good stability, such as an oven-controlled oscillator, or connect a more accurate external reference, such as a rubidium standard or a clock disciplined by a satellite navigation system.
- Allow the instrument and the oscillator to stabilize thermally before measuring.
- Choose a gate time in line with the required resolution. Long gates average out noise but do not allow rapid variations to be seen.
- Use a reciprocal counter for low-frequency signals, which maintains high resolution with short gates.
- Ensure an adequate signal level at the input and good coupling, to avoid false triggering from noise.
Measuring an oscillator and comparing it
To evaluate the error of an oscillator, what matters is measuring it against a better reference. If all the instruments on a test bench share their references, time base errors largely cancel, and spectrum analyzers and generators also gain accuracy.
What a counter does not measure
It does not distinguish between signals. If there is more than one component, it may count the strongest one or a mixture, so filtering is advisable. Nor does it see phase noise, which requires other methods.