How is a TRL calibration performed and when is it used?

A direct connection, a reflect and one or more lines of known length are measured; since only the line needs a precise definition, the method is highly accurate in media where good standards do not exist, such as waveguide, microstrip or wafers.

The idea

TRL stands for Thru, Reflect, Line. Unlike SOLT, it does not require short, open and load standards whose value must be known accurately. With three types of measurements the same system errors are determined, and accuracy rests on a few assumptions.

What is measured

  • Thru. A zero-length or very short connection between the reference planes, which defines those planes.
  • Reflect. A reflector with a high reflection coefficient at each port, such as a short or an open. It need not be known accurately, but it must be the same at both ports and its phase must be known approximately.
  • Line. A transmission line of the same impedance as the thru and of greater length, which must be known.

What you get from it

The characteristic impedance of the line becomes the reference impedance of the calibration. The quality of that line therefore sets the accuracy of all measurements. This is an advantage when a precise line is manufactured, and a limitation when it cannot be made uniform.

Limitations

  • The electrical length difference between the line and the thru must be far from multiples of half a wavelength. Near those points the method becomes ill-conditioned. A single line therefore covers a band of frequencies with a limited ratio between its ends, and several lines are used for wider bands.
  • At low frequencies, the required line would be too long.
  • Calibration structures manufactured in the same medium as the device are needed.

When it is used

In on-wafer measurements, in printed circuit board fixtures, in waveguide, and in any environment where manufacturing a line standard is easier than manufacturing a quality load or open. It also serves as a reference method for verifying other standards.

How to interpret the result

After calibration, the measurement plane is at the center of the thru, or at its ends if it is defined with a correction. A verification with a known device confirms that the reflector, the lines and the contacts worked.