How is a time-domain measurement made with a VNA?

The frequency response is measured and the analyzer converts it to the time domain with an inverse transform; the result shows where each reflection is and how large it is along a line or device.

What the instrument does

A vector network analyzer measures, step by step, magnitude and phase at a set of frequencies. Applying an inverse Fourier transform to this complex response gives what would be seen if the device were excited with an impulse or a step. The horizontal axis becomes time, or distance when multiplied by the propagation velocity of the cable.

What is seen

With a reflection measurement, each impedance discontinuity, such as a connector, a splice, damage or an open end, appears as a peak at the corresponding position. The shape of the change identifies where the defect is and whether the section has a higher or lower impedance.

Parameters that set what can be achieved

  • Resolution. It depends on the measurement bandwidth: the wider the frequency range, the more finely two nearby reflections are separated.
  • Alias-free range. It depends on the frequency step: a smaller step makes it possible to see farther before the responses repeat. Seeing farther requires using more points.
  • Low-pass mode. It requires harmonically related frequencies starting near zero; it supports impulse and step responses and makes it possible to see the impedance type of the discontinuity.
  • Bandpass mode. It works with any frequency range, but gives only the envelope of the impulse response.

Windows and gating

Limiting the measurement to a finite range generates sidelobes that hide small reflections near large ones. A smoothing window reduces them, at the cost of somewhat poorer resolution. Time gating makes it possible to isolate a reflection, discard the rest and return to the frequency domain to see the response of that element alone.

What to consider

The propagation velocity of the cable must be entered correctly for the distance to be right. Vector calibration is essential. And the interpretation assumes simple reflections: multiple reflections between discontinuities can appear as false elements farther away.