24/7 emergencies 814 777 6921812 230 6079WhatsApp 811 786 0768
Electrónica MonterreyElectrónica MonterreyLaboratorios Creativeit Talk to an engineer
Menu

How does an electrochemical gas sensor work?

An electrochemical gas sensor works like a small cell: the gas passes through a membrane, reacts at a working electrode and the current generated is proportional to its concentration.

Principle

The target gas enters the sensor through an opening with a porous membrane that regulates how much passes and blocks water and dust. Inside, at the working electrode, the gas is oxidized or reduced in the presence of a catalyst. This reaction releases or consumes electrons and produces an electric current. The more gas that arrives, the higher the current. A circuit measures it and converts it into a concentration.

Components

  • Membrane and capillary: control the diffusion of gas toward the electrode.
  • Working electrode: where the reaction with the gas of interest takes place.
  • Counter electrode: completes the circuit with the opposite reaction.
  • Reference electrode (in many designs): keeps the working electrode potential stable.
  • Electrolyte: a liquid or gel medium that allows ions to pass between the electrodes.

Why it is selective, and how far that goes

Selectivity is achieved through the catalyst, the electrolyte, the applied potential and internal filters. Even so, no sensor is perfect: it may respond to other similar gases. This phenomenon is called cross-sensitivity. Data sheets indicate which gases interfere and by how much.

Oxygen sensor: a special case

The oxygen cell is usually galvanic: oxygen reacts with an electrode that is consumed, and the cell generates the current by itself. That is why its life is limited and depends on exposure to oxygen.

What influences the reading

Factor Effect
Temperature Changes the reaction rate and diffusion; the instrument partly compensates.
Humidity Prolonged drying or excess water alters the electrolyte and the response.
Aging Gradual loss of sensitivity and zero drift.
High exposure Very high concentrations can saturate or damage the sensor.
Contamination Dirt or substances on the membrane reduce the passage of gas.

What the engineer sees in practice

A healthy sensor responds quickly and returns to zero in clean air. An old sensor responds slowly, takes time to return or shows an unstable zero. That is why the functional test with gas and periodic calibration are part of use, and sensors are replaced according to the manufacturer’s instructions. Use in safety applications must be in the hands of trained personnel, in accordance with applicable regulations.

This guide is for information only. Work involving electrical hazards, heights or hazardous atmospheres must be carried out by qualified, authorized personnel in accordance with applicable regulations.

We do real tests, on video! Watch them on our YouTube channel →

Do you have a problem nobody has been able to measure?

Talk directly to the engineer in charge, with no middlemen.

Talk to an engineer