What sets it apart from a spectrum analyzer
A conventional spectrum analyzer measures amplitude versus frequency and discards the phase. A vector signal analyzer preserves amplitude and phase, that is, the complex signal. With both, the modulation can be reconstructed and the behavior of the transmitter evaluated in detail.
Signal chain
- Input conditioning: attenuation and filtering that protect the converter input and adjust the level to it.
- Frequency conversion: a local oscillator and mixers move the signal to an intermediate frequency or to baseband.
- Digitization: an analog-to-digital converter samples the signal. In the in-phase and quadrature representation, the signal is described by two components, I and Q.
- Digital processing: filtering, carrier and symbol clock synchronization, and demodulation, according to the format chosen by the user.
- Display: spectrum, constellation, eye diagram, error vector magnitude, phase and amplitude errors, and symbol tables.
What is obtained
- Modulation quality, through the error vector magnitude and the shape of the constellation.
- Diagnosis of impairments: imbalance between I and Q, local oscillator leakage, phase noise, non-linearity.
- Time-domain measurements of turn-on transients, bursts and frequency hops.
- Channel power and adjacent channel power.
Conditions for a valid measurement
The analyzer must be configured with the same format as the transmitter: modulation type, symbol rate, pulse-shaping filter and synchronization sequence. The analyzer also introduces its own errors. If the instrument has a residual EVM comparable to that of the transmitter, it is impossible to separate how much of the error belongs to each. The input level must be set to make use of the dynamic range without saturating the converter.
How to interpret what is seen
A round cloud of points suggests noise, points forming arcs suggest phase noise, and a constellation shifted off center suggests carrier leakage. This visual reading guides the correction before looking at figures.