The signal chain
- Input attenuator. Adjusts the level to protect the mixer and avoid distortion.
- Input filter or preselector. Limits the band and reduces unwanted responses, such as the image frequency.
- Mixer and swept local oscillator. The local oscillator varies its frequency along a ramp. Mixing translates the part of the input spectrum that corresponds to each instant to a fixed intermediate frequency.
- Intermediate-frequency filter. This is the resolution filter: it passes only what falls around it. Its width determines how two nearby signals appear.
- Envelope detector. Converts the filtered signal into a level.
- Video filter. Averages the detected level and smooths the noise on the trace.
- Display. The ramp that sweeps the local oscillator also drives the horizontal axis, so each position corresponds to a frequency.
What the principle implies
- At each instant the instrument looks at only a narrow band. A brief event that occurs outside the band being observed at that moment is not detected. Intermittent signals may not appear, or may appear at varying levels from sweep to sweep.
- The sweep cannot be arbitrarily fast: the narrow filter needs time to settle, which ties the resolution bandwidth to the sweep time.
- The on-screen level is relative to the instrument’s calibration, and its level accuracy is not that of a power meter.
Measurement limitations
The mixer generates its own distortion products if the input signal is very strong; to check whether distortion is real, add attenuation and see whether it changes. The instrument’s noise floor sets the smallest visible signal, and it drops with a narrower filter.
When another type of analyzer is the better choice
For transient events or frequency-hopping signals, an analyzer based on the fast Fourier transform or a real-time analyzer captures the entire interval instead of a sequential sweep.