Destructive or nondestructive failure analysis?

Destructive analysis opens, cuts or alters the part to see inside it; nondestructive analysis examines it without damaging it. Start with nondestructive analysis and move to destructive analysis only when there is nothing more to learn without it.

The practical difference

A nondestructive analysis leaves the part intact and repeatable: visual inspection, photography, radiography, ultrasound, thermography, electrical measurements. A destructive analysis changes it irreversibly: sectioning, decapsulating an integrated circuit, making metallographic cuts or taking it to rupture.

Why the order matters

Evidence can only be destroyed once. If a part is cut before its condition is documented, information is lost that will not come back, such as the exact location of a fracture or the position of a residue. That is why the rule is to move from the least invasive to the most invasive.

Recommended sequence

  1. Photograph and describe the condition as received.
  2. Carry out external inspections and measurements.
  3. Apply internal nondestructive techniques (radiography, ultrasound, tomography).
  4. Plan the destructive analysis: what is to be confirmed and where to cut.
  5. Execute with care so as not to introduce artifacts, and document each step.

When destructive analysis is justified

  • When external techniques cannot distinguish between hypotheses.
  • To observe the failure mechanism in the material: microstructure, fracture, internal corrosion.
  • To verify a manufacturing defect inside a component.

How to interpret the result

Destructive analysis must distinguish damage from the failure from damage caused while preparing the sample. A poorly made cut can create cracks that look like original defects. That is why healthy reference parts are used, when they exist, and compared.

Additional considerations

If the part is the only sample, or there is contractual or legal liability over the cause, it is advisable to agree on the plan with the interested parties before intervening. When several parts have the same failure, one can be kept untouched as a control.