24/7 emergencies 814 777 6921812 230 6079WhatsApp 811 786 0768
Electrónica MonterreyElectrónica MonterreyLaboratorios Creativeit Talk to an engineer
Menu

Attenuating probe or unity-ratio probe?

An attenuating probe reduces the signal by a factor, usually ten, and in exchange increases input resistance, reduces capacitive loading and extends the voltage range; a unity-ratio probe does not attenuate, so it reads small signals with a better signal-to-noise ratio, but it loads the circuit more and has less bandwidth.

What changes between the two

A unity-ratio probe connects the tip to the oscilloscope input through a cable, with no divider. An attenuating probe includes a compensated resistive divider at the tip. The difference is not just the scale: it changes what the circuit sees when the probe is connected and what the instrument receives.

Practical comparison

Aspect Attenuating Unity ratio
Resistive loading Higher input resistance That of the channel itself
Capacitive loading Lower Higher, due to the cable and the input
Bandwidth Higher Lower
Voltage range Greater, due to the attenuation Smaller
Small signals Worse signal-to-noise ratio Better, not attenuated

When each one makes sense

  • Attenuating: for general use, fast signals, load-sensitive circuits and moderate voltage levels.
  • Unity ratio: for low-frequency, low-amplitude signals, such as power supply ripple or a sensor output, where the instrument’s own noise would weigh too heavily if the signal were attenuated.

Many probes have a switch between the two positions. In the unity position, bandwidth drops markedly and capacitive loading increases, so it should not be used for fast signals.

What to check first

  1. The attenuation factor set on the channel must match the actual switch position. If it does not, amplitude readings end up multiplied or divided by that factor.
  2. Compensation must be adjusted with the probe in its attenuating position, because the unity ratio has no compensation.
  3. The probe’s voltage limits change with the position, and the lower limit is the one that matters.

If you see the signal at twice or one tenth of the expected amplitude, check that factor first before suspecting the circuit.

We do real tests, on video! Watch them on our YouTube channel →

Do you have a problem nobody has been able to measure?

Talk directly to the engineer in charge, with no middlemen.

Talk to an engineer