24/7 emergencies 814 777 6921812 230 6079WhatsApp 811 786 0768
Electrónica MonterreyElectrónica MonterreyLaboratorios Creativeit Talk to an engineer
Menu

How is the permittivity of an insulating liquid measured?

The relative permittivity of an insulating liquid is measured in a guarded electrode cell: the capacitance with the liquid is determined and divided by the capacitance of the same cell empty or with a reference fluid, at controlled temperature and frequency.

Principle

Relative permittivity is the ratio between the capacitance of a capacitor filled with the material and the capacitance it would have with a vacuum between the same electrodes. With a cell of fixed geometry, it is enough to measure the capacitance with the liquid and relate it to the capacitance of the empty cell. In practice, the capacitance with air is used as an approximation of vacuum, or the cell is calibrated with a liquid of known permittivity.

Equipment

  • Test cell: concentric or flat electrodes, made of materials inert to the liquid, with a guard electrode to eliminate the edge effect.
  • Capacitance meter: a bridge or LCR analyzer that also measures the dissipation factor at the chosen frequency.
  • Temperature control: a bath or chamber, because permittivity depends on temperature.

General procedure

  1. Clean and dry the cell and measure its empty capacitance, recording the temperature.
  2. Fill the cell with the liquid, avoiding bubbles, and wait for thermal equilibrium.
  3. Measure capacitance and dissipation factor at the defined frequency and voltage.
  4. Calculate the permittivity as the ratio of capacitances, subtracting the stray capacitances of cables or accessories that do not belong to the cell.
  5. Repeat and report the measurement temperature, frequency and voltage.

Sources of error

  • Moisture and particles: these alter the permittivity and, above all, the losses.
  • Stray capacitances from wiring and connections.
  • Temperature gradients within the liquid.
  • Air bubbles trapped between electrodes.
  • Electrode polarization at low frequencies in conductive liquids.

How to interpret the result

A permittivity value without its measurement temperature and frequency has no comparative meaning. In mineral insulating oils the permittivity is relatively low and varies little between healthy samples; changes due to contamination or aging show up better in the dissipation factor and resistivity, which are usually measured in the same cell. The standard IEC 60247 covers these measurements on insulating liquids; its current edition should be consulted.

Warning: if the cell is energized at high voltage, the work must be done by qualified personnel, following the applicable safety procedures.

We do real tests, on video! Watch them on our YouTube channel →

Do you have a problem nobody has been able to measure?

Talk directly to the engineer in charge, with no middlemen.

Talk to an engineer