24/7 emergencies 814 777 6921812 230 6079WhatsApp 811 786 0768
Electrónica MonterreyElectrónica MonterreyLaboratorios Creativeit Talk to an engineer
Menu

How is a digital multimeter calibrated in general terms?

A digital multimeter is calibrated by applying known signals, generated by a traceable calibrator or standard, on each selected function and range, and comparing the multimeter reading with the reference value to obtain the error and its uncertainty.

Preconditions

Before measuring, the equipment is inspected (leads, probes, display, battery or power supply), allowed to stabilize at a controlled room temperature, and given the warm-up time indicated by the manufacturer. The standards must have current traceability and an uncertainty clearly lower than the tolerance of the instrument being calibrated.

General sequence

  1. Plan the points: the functions (DC and AC voltage, current, resistance, sometimes frequency or capacitance), the ranges and several values per range, including values near full scale and near zero. For AC, different frequencies are included.
  2. Verify zero: for DC voltage with the input short-circuited, and for resistance by compensating for lead resistance or using a four-wire connection when resistances are low.
  3. Apply the reference signal with the calibrator, wait for it to stabilize and record several multimeter readings.
  4. Calculate the error: multimeter reading minus the standard’s value at each point.
  5. Estimate the uncertainty with the contributions of the standard, resolution, repeatability, temperature effect and connections.
  6. Evaluate the result against the specification or agreed criterion, applying the defined decision rule.

Calibrating is not adjusting

Calibrating means comparing and documenting. Adjusting means modifying the instrument to reduce its error. If an adjustment is made, as-found and as-left data are reported, because the as-found data show whether the instrument drifted in use.

How to interpret the result

Errors are compared with the manufacturer’s specification as of the calibration date and with the uncertainty of the process itself. An error within the limit but with a large uncertainty is less informative than a similar one with a small uncertainty. Over time, the trend in errors between calibrations helps decide the appropriate interval.

What an engineer sees in practice

The certificate presents a table by function, range and point, with the nominal value, reading, error and uncertainty. The most frequent differences appear in the high AC ranges and at high resistances, where wiring and environmental conditions weigh more.

Warning: the high voltages and currents applied during calibration present an electrical hazard. The work must be performed by qualified personnel, following applicable regulations and safety procedures.

We do real tests, on video! Watch them on our YouTube channel →

Do you have a problem nobody has been able to measure?

Talk directly to the engineer in charge, with no middlemen.

Talk to an engineer