Basis
A noise source generates noise in two states: off, with an effective temperature close to ambient, and on, with a much higher effective temperature. The difference is characterized by a calibrated value, the excess noise ratio or ENR, which depends on frequency. If the noise power at the output of the device under test is measured in both states, the ratio of the power with the source on to the power with it off is the Y factor.
Calculation
The total noise at the output is the source noise amplified by the device, plus the noise the device itself adds. Turning the source on changes only the first term. This relationship makes it possible to solve for the noise factor: it is the excess noise of the source, in linear form, divided by the Y factor minus one. The noise figure is obtained by converting the result to decibels. The gain of the device is calculated from the same measurement and does not need to be known in advance.
Procedure
- Connect the noise source to the input of the device and a noise receiver, such as a spectrum analyzer or noise figure meter, to the output.
- Calibrate the system without the device to measure the receiver’s own noise.
- Insert the device and measure the powers with the source off and on, at each frequency of interest.
- Correct for the receiver’s contribution using the cascade formula, since its noise is not negligible when the device gain is low.
Sources of error
- Uncertainty of the source ENR and its variation with frequency.
- Impedance mismatch between the source, the device and the receiver, which modifies the effective noise and gain.
- Low device gain, which lets the receiver dominate and makes the correction large.
- External noise or spurious signals and an inappropriate measurement bandwidth.
- Cable and adapter losses ahead of the device, which add directly to the noise figure.
- Temperature different from that assumed for the off state.
What to check first
Before interpreting the result, verify that the receiver-only measurement is stable and repeatable, and that the auxiliary elements have a known loss. The measurement is usually compared with a known reference device to check the setup.