How is the resistivity of an insulating oil measured?

The resistivity of an insulating oil is measured by applying a DC voltage between the electrodes of a cell of known geometry at a controlled temperature, measuring the current after a fixed electrification time, and calculating the resistance using the geometric constant of the cell.

What it measures

Volume resistivity indicates how poorly the oil conducts, that is, how many free ions it contains. A clean oil has very high resistivity; polar contaminants, water, particles and aging products reduce it. It is the direct-current counterpart of the dissipation factor, which measures loss under alternating current.

Measurement principle

  1. A test cell with concentric electrodes, including a guard electrode that eliminates surface currents, is filled and the sample is brought to the specified temperature.
  2. A stable DC voltage is applied.
  3. After a defined time, the current is measured. The wait is needed because charging and polarization currents flow at first that do not correspond to true conduction.
  4. The resistance is obtained by dividing the voltage by the current, and the resistivity by multiplying that resistance by the geometric factor of the cell, which relates the area of the electrodes to their spacing.

Because the currents are extremely small, a high-impedance electrometer and shielding against interference are needed.

Factors that affect the result

  • Temperature: resistivity decreases sharply on heating. A small temperature variation produces a significant error, so thermal control is essential.
  • Cell cleanliness: any residue of solvent or of previous samples contaminates the measurement.
  • Electrification time and polarity: measuring too early gives low values because of the polarization current. Some methods average both polarities.
  • Ambient humidity and handling: the oil can absorb water during filling.
  • Field intensity: if it is very high, conduction ceases to be ohmic.

How to interpret the result

The value is compared with that of new oil and with the criteria of the method for the type of equipment and, above all, its trend over time is observed. A fall in resistivity is usually accompanied by a rise in the dissipation factor. If only one of the two changes, the measurement should be reviewed.

When to distrust a value

A very low result on a sample that appears clean makes it necessary to verify the cleanliness of the cell and the temperature before concluding that the oil is degraded.