From radiation to image
Every body above absolute zero emits electromagnetic radiation, and the amount rises rapidly with temperature (according to the Stefan-Boltzmann law, with the fourth power of absolute temperature). At ordinary temperatures the emission is mainly infrared, invisible to the eye.
Stages inside the camera
- Optics: ordinary glass does not transmit thermal infrared well, so lenses are made from materials such as germanium. They focus the radiation onto the focal plane.
- Detector: an array of sensitive elements. Each one receives radiation from a small portion of the scene and generates an electrical signal that depends on the power received.
- Processing: differences between elements are corrected, the calibration is applied and the signal is converted to temperature.
- Display: each level is assigned a color or shade of gray using a palette. The colors are a convention, not a physical property.
What the camera actually receives
The signal does not come from the object alone. It adds up three contributions: what the object emits, what it reflects from its surroundings, and what the atmosphere adds or attenuates along the path. To isolate the object’s emission, the software needs the surface emissivity, the reflected apparent temperature, the distance and the air conditions.
What an engineer sees in practice
An image in which differences in brightness represent differences in radiation, not necessarily in temperature. A shiny surface can look cold when it is hot, because it emits little and reflects its surroundings. That is why the measurement settings carry as much weight as the focus.
How to interpret the result
Treat the image as a tool for comparing similar elements, always verify the measurement parameters, and confirm with another method when the decision is critical.